Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: AEMC 3900: (1 year) CAL VER /5520 DATE: 18-Feb-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 79 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 DISP {Procedure not formally tested with UUT.} 1.002 DISP {See procedure README file.} 1.003 ASK- R P F V 1.004 ASK+ K X 1.005 HEAD PRELIMINARY INSTRUCTIONS 1.006 DISP Connect UUT to AC power and turn unit ON. 1.006 DISP 1.006 DISP Warm-up time: 30min - REQUIRED 1.006 DISP Temperature: 23 +/-2 degC 1.006 DISP Humidity: 55 +/-10% 1.007 DISP [32] WARNING 1.007 DISP HIGH VOLTAGE is used or exposed during the performance 1.007 DISP of this calibration. DEATH ON CONTACT may result if 1.007 DISP personnel fail to observe safety precautions. 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT V Input 1.008 DISP [32] 5520A NORMAL LO to UUT N Input(neutral) 1.009 HEAD {VOLTAGE} 1.010 DISP Select Volts. 1.010 DISP Select the 200V Range. 1.011 HEAD VOLTAGE : {200V Range} 1.012 5520 200 180.0V 0.4/ 60H SI 2W 2.001 5520 200 180.0V 0.4/ 0.1U 500H SI 2W 3.001 HEAD VOLTAGE : {600V Range} 3.002 DISP Select the 600V Range. 3.003 5520 600 580.0V 0.4/ 60H SI 2W 4.001 HEAD {CURRENT} 4.002 DISP Leave voltage connections in place! 4.003 DISP Connect the standard 1000:1 clamp-on current probe 4.003 DISP to the 3900 current input jacks. 4.003 DISP 4.003 DISP Connect a Fluke 5520A/COIL to the 5520A 20A and AUX LO. 4.004 DISP Clamp the current clamp around the 50-turn coil. 4.005 DISP Select Amps. 4.005 DISP Select the 200A Range. 4.006 5520 1000 180A 1/ 3U 60H SI HC LT5 5.001 DISP Select the 2000A Range. 5.002 5520 1000 500A 1/ 3U 60H SI HC LT5 6.001 HEAD {POWER FACTOR TESTS} 6.002 DISP Select COS. 6.002 DISP Select the 2000A Range. 6.003 M5520 200V -90deg SI 6.004 5520 400A 60H SI HC S LT5 6.005 ACC 000pf 0.086U 6.006 MEMI Enter the UUT Power Factor reading. 6.007 MEME 6.008 MEMC pf 11U 7.001 END